Question: A portable photometer with a linear response to radiation registered 75.5 (A with a blank solution in the light path. Replacement of the blank with
(a) The percent transmittance of the sample solution.
(b) The absorbance of the sample solution.
(c) The transmittance to be expected for a solution in which the concentration of the absorber is one third that of the original sample solution.
(d) The transmittance to be expected for a solution that has twice the concentration of the sample solution.