Question: A spectrometer operates at 800 MHz for proton, and it is desired to record a spectrum covering a shift range of 15 ppm. Assuming that
A spectrometer operates at 800 MHz for proton, and it is desired to record a spectrum covering a shift range of 15 ppm. Assuming that the receiver reference frequency is placed in the middle of this range, whqt range of frequencies (in Hz) is covered by the spectrum, and what would the sampling interval (dwell time) have to be?
Step by Step Solution
3.45 Rating (155 Votes )
There are 3 Steps involved in it
15 ppm at 800 MHz is 15 800 12 000 Hz The range of ... View full answer
Get step-by-step solutions from verified subject matter experts