Question: Detection limit. A sensitive chromatographic method was developed to measure sub-part per-billion levels of the disinfectant by-products iodate (IO 3 ), chlorite (ClO 2 ),

Detection limit. A sensitive chromatographic method was developed to measure sub-part per-billion levels of the disinfectant by-products iodate (IO 3 ), chlorite (ClO 2 ), and bromate (BrO 3 ) in drinking water. As the oxyhalides emerge from the column, they
react with Br to make Br 3, which is measured by its strong absorption at 267 nm. For example, each mole of bromate makes 3 mol of Br 3 by the reaction
BrO3- + 8Br- + 6H+ †’ 3Br3- + 3H2O
Bromate near its detection limit gave the following chromatographic peak heights and standard deviations (s). For each concentration, estimate the detection limit. Find the mean detection limit. The blank is 0 because chromatographic peak height is measured from the baseline adjacent to the peak. Because blank 0, relative standard deviation applies to both peak height and concentration, which are proportional to each other. Detection limit is 3s for peak height or concentration.
Detection limit. A sensitive chromatographic method was developed to measure

Bromate Peak height Relative Number of measurements concentration (arbitrary (Hg/L) 0.2 0.5 1.0 2.0 units) 17 31 56 standard deviation(%) 14.4 6.8 3.2 1.9

Step by Step Solution

3.32 Rating (176 Votes )

There are 3 Steps involved in it

1 Expert Approved Answer
Step: 1 Unlock

For a concentration of 02 jagL the relative standard deviation of 144 correspond... View full answer

blur-text-image
Question Has Been Solved by an Expert!

Get step-by-step solutions from verified subject matter experts

Step: 2 Unlock
Step: 3 Unlock

Document Format (1 attachment)

Word file Icon

878-E-C-E-E-C (1724).docx

120 KBs Word File